The Omron M3 Blood Pressure Monitor is an automatic upper-arm blood pressure monitor with IntelliSense technology for comfortable inflation, a large LCD display with backlight, hypertension indicator, irregular heartbeat detection, body movement detection, memory storage for two users (up to 60 readings each), and WHO classification indicator. It is clinically validated and designed for easy home use. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit with LCD display, Arm cuff with D-ring, AC adapter, Storage case, Instruction manual.
| Feature | Description |
|---|---|
| IntelliSense Technology | Automatically inflates to optimal pressure for comfortable measurement |
| Large LCD with Backlight | Easy-to-read display, visible in low light |
| Hypertension Indicator | Displays symbol if systolic/diastolic readings are high |
| Irregular Heartbeat Detection | Detects and displays symbol for irregular pulse during measurement |
| Body Movement Detection | Alerts if excessive movement is detected |
| Dual User Memory | Stores up to 60 readings for each of two users |
| WHO Classification Indicator | Shows color-coded bar graph based on WHO guidelines |
| Date & Time Stamp | Records date and time of each measurement |
| Average Reading | Calculates average of last three readings (morning or evening) |
| Automatic Power Off | Conserves battery life |
Read all instructions before use. For home use only.
WARNING! This device is not a substitute for professional medical diagnosis. Always consult your doctor about your blood pressure.
Unpack and verify all items are present.
CAUTION! Do not use if any part is damaged. Contact customer service.
Front Panel: START/STOP button, MEMORY button, USER 1/USER 2 buttons, LCD display. Cuff: Air tube connector, D-ring, artery mark. Rear: Battery compartment, AC adapter jack.
LCD Icons: User indicator, Memory full, Heartbeat symbol, Movement symbol, Hypertension indicator, WHO bar graph, Battery low, Date/Time, Systolic/Diastolic/Pulse readings.
Prepare the device for initial use.
Tip: Rest for 5 minutes before first measurement. Avoid caffeine, smoking, and exercise for 30 minutes prior.
Follow these steps for accurate readings.
WARNING! Do not inflate cuff without it being wrapped around your arm.
Refer to the WHO classification indicator on the display.
| Category | Systolic (mmHg) | Diastolic (mmHg) | Indicator Color |
|---|---|---|---|
| Optimal | < 120 | < 80 | Green |
| Normal | 120-129 | 80-84 | Light Green |
| High Normal | 130-139 | 85-89 | Yellow |
| Hypertension Grade 1 | 140-159 | 90-99 | Orange |
| Hypertension Grade 2 | 160-179 | 100-109 | Red |
| Hypertension Grade 3 | ≥ 180 | ≥ 110 | Flashing Red |
Note: The hypertension indicator ( ) will appear if systolic ≥ 135 or diastolic ≥ 85.
Stores readings for two users separately.
Tip: Take measurements at the same time each day for consistent tracking.
Unplug or remove batteries before cleaning.
CAUTION! Do not immerse any part in water. Do not expose to extreme temperatures or humidity.
| Display | Meaning | Action |
|---|---|---|
| E-1 | Cuff not properly attached or air leakage | Reattach cuff securely; check tube connection |
| E-2 | Excessive movement during measurement | Remain still and retake measurement |
| E-3 | Pulse too irregular or weak | Rest and try again; consult doctor if persistent |
| E-4 | Pressure over 300 mmHg | Remove cuff, restart device |
| E-5 | Device malfunction | Contact customer service |
| Lo | Low batteries | Replace all batteries |
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries, improper insertion | Replace batteries; ensure correct polarity. |
| Inaccurate reading | Incorrect cuff position, movement | Reposition cuff correctly; remain still and retake. |
| Cuff does not inflate | Tube not connected, battery low | Check tube connection; replace batteries. |
| Memory not working | Memory full, button error | Clear memory; ensure correct user button is pressed. |
| Error codes appear | See Error Messages section | Follow corrective action for specific code. |
Reset: Remove batteries and AC adapter for 1 minute, then reinsert.
Service: For technical support, visit www.omron-healthcare.com or contact your local Omron representative.