The Omron BP786N is an advanced upper arm blood pressure monitor with IntelliWrap™ Cuff, TruRead™ technology, and Bluetooth® connectivity. It features a wide-range D-ring cuff (22-42 cm), dual LCD screens, irregular heartbeat detection, body movement detection, hypertension indicator, 200-user memory with date/time stamp, averaging of last 3 readings, AC adapter compatibility, and battery operation. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit, IntelliWrap™ Cuff (22-42 cm), AC adapter, storage case, instruction manual, quick start guide.
| Feature | Description |
|---|---|
| IntelliWrap™ Cuff | Pre-formed cuff with 360° correct positioning indicator |
| TruRead™ Technology | Takes 3 consecutive readings and provides an average |
| Bluetooth® Connectivity | Syncs data with Omron Connect app (iOS/Android) |
| Dual LCD Display | Shows systolic/diastolic pressure, pulse, and date/time |
| Irregular Heartbeat Detection | Detects and displays irregular heartbeat symbol |
| Body Movement Detection | Alerts if movement is detected during measurement |
| Hypertension Indicator | Displays color-coded indicator based on ESH guidelines |
| 200-User Memory | Stores up to 200 readings with date/time stamp |
| AC Adapter Port | For use with included AC adapter (battery backup) |
| START/STOP Button | Initiates and stops blood pressure measurement |
Read all instructions before use. For home use only.
WARNING! Not for treatment or diagnosis of disease. Measurements are for reference only.
Unpack and verify all items are present.
CAUTION! Do not use if package is damaged or items are missing.
Front Panel: START/STOP button, Memory button, Clock button, Bluetooth® button, Dual LCD display. Cuff: IntelliWrap™ design with metal D-ring, air tube connector, 360° positioning guide.
Display Symbols: Systolic/Diastolic pressure, Pulse rate, Irregular heartbeat, Body movement, Hypertension indicator, Bluetooth® connection, Low battery, Memory index, Date/Time.
Buttons: START/STOP (start/stop measurement), MEMORY (review stored readings), CLOCK (set date/time), BLUETOOTH (sync with app).
Prepare device for first use.
Tip: Use AC adapter for extended use to conserve batteries.
For accurate readings, follow these steps.
WARNING! Do not inflate cuff over 299 mmHg. Stop if pain occurs.
Sync data for tracking and sharing.
Note: App requires iOS 10.0+ or Android 5.0+.
Store and review up to 200 readings.
Tip: Sync regularly with app to free device memory.
Keep device clean and functioning properly.
CAUTION! Do not expose to direct sunlight, moisture, or dust.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace batteries; ensure correct polarity. |
| Error code E1 | Cuff not attached | Connect cuff properly to main unit. |
| Error code E2 | Cuff over-inflated | Deflate cuff and restart measurement. |
| Irregular heartbeat symbol | Arrhythmia detected | Consult physician if symbol appears frequently. |
| Bluetooth® not connecting | Distance/interference | Keep within 3 ft; ensure Bluetooth® is enabled on phone. |
| Inaccurate readings | Improper cuff position | Reposition cuff at heart level; ensure proper fit. |
Reset: Remove batteries for 5 minutes to reset device.
Service: 1-800-634-4350; www.omronhealthcare.com