Omron Blood Pressure Monitor Manual BP652

The Omron BP652 is a fully automatic upper arm blood pressure monitor with IntelliSense technology for comfortable inflation, a wide-range D-Ring cuff (22-42 cm), irregular heartbeat detection, hypertension indicator, 90-reading memory with date and time stamp, averaging of last 3 readings, and body movement detection. It features a large LCD display, one-touch operation, and is clinically validated for accuracy. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.

1. Features at a Glance 2. Safety Information 3. Parts Identification 4. Specifications 5. Initial Setup 6. Taking a Measurement 7. Understanding the Display 8. Memory Function 9. Care & Maintenance 10. Regulatory Information 11. Troubleshooting

Features at a Glance

Key components: Main unit with LCD display, D-Ring Cuff, AC Adapter, Instruction Manual, Storage Case.

FeatureDescription
IntelliSense TechnologyAutomatically inflates to the optimal level for comfortable measurement
D-Ring Cuff (22-42 cm)Pre-formed, easy-to-wrap cuff for proper fit on the upper arm
Irregular Heartbeat DetectionDetects and displays an icon if an irregular heartbeat is detected during measurement
Hypertension IndicatorDisplays a symbol if systolic/diastolic readings are above WHO guidelines
90-Reading MemoryStores up to 90 readings with date and time stamp
Average of Last 3 ReadingsCalculates and displays the average of the last 3 readings taken within 10 minutes
Body Movement DetectionAlerts user if excessive movement is detected during measurement
One-Touch OperationSimple START/STOP button for easy use
Large LCD DisplayEasy-to-read backlit display with clear icons
Power Options4 x AA batteries or AC adapter (included)

Safety Information

Read all instructions before use. For personal home use only.

  1. Consult your physician: Before using if you have arrhythmia, other heart conditions, or are pregnant.
  2. Do not use on infants, children, or persons unable to express themselves.
  3. Do not take measurements on an arm with an intravenous drip or surgical wound.
  4. Wait at least 1 hour after exercising, bathing, eating, or drinking caffeine/alcohol before measuring.
  5. Remain still and quiet during measurement; do not talk.

WARNING! This device is not a treatment device. Measurements are for reference only. Always consult a physician for diagnosis and treatment.

Parts Identification

Specifications

Initial Setup

Before first use, install batteries or connect AC adapter, set date and time.

  1. Install Batteries: Open battery compartment, insert 4 AA batteries observing polarity, close cover.
  2. Set Date and Time: Press the MEMORY button for more than 3 seconds until the year flashes. Use START/STOP to adjust value, press MEMORY to confirm and move to month, day, hour, minute. Settings complete automatically.
  3. Connect Cuff: Plug the cuff air tube firmly into the air plug on the left side of the main unit.

Note: If using AC adapter, plug it into the jack on the right side of the unit before connecting to power outlet. Using the AC adapter disables battery power.

Taking a Measurement

For accurate results, measure at the same time each day, in a quiet place, while seated.

  1. Prepare: Rest for 5 minutes before measuring. Avoid caffeine, tobacco, and exercise for 30 minutes prior.
  2. Apply Cuff: Sit on a chair with feet flat on floor. Remove tight clothing from left upper arm. Pass end of cuff through metal D-Ring to form a loop. Place cuff on bare arm 1-2 cm above elbow with air tube centered on inner arm and pointing down. Secure snugly.
  3. Position Arm: Rest arm on a table so cuff is at heart level. Palm should be up.
  4. Start Measurement: Press the START/STOP button. The cuff will inflate automatically. Remain still and quiet until measurement is complete.
  5. Read Results: Systolic, diastolic pressure, and pulse rate will display. Note any warning icons (irregular heartbeat, body movement).
  6. Power Off: Unit turns off automatically about 1 minute after measurement.

Tip: For consistent results, take 2-3 measurements, waiting 2-3 minutes between each. The unit will average the last 3 readings taken within 10 minutes if this feature is activated.

Understanding the Display

The LCD shows various icons and values:

SymbolMeaning
SYS / DIASystolic / Diastolic pressure (mmHg)
Pulse SymbolPulse rate (beats per minute)
Irregular Heartbeat IconDetected irregular rhythm during measurement
Hypertension IndicatorSystolic ≥135 mmHg or Diastolic ≥85 mmHg (based on WHO)
Body Movement IconExcessive movement detected; retake measurement
Memory Index (e.g., 01)Current memory position (1-90)
Date & TimeStamp of when reading was taken
Average IconDisplayed when showing average of last 3 readings
Battery IconIndicates battery level (replace when low)
AC Adapter IconDisplayed when using AC power

Memory Function

The device stores up to 90 readings with date and time.

  1. View Memory: Press the MEMORY button briefly. The most recent reading displays with its memory index. Press MEMORY repeatedly to scroll through older readings.
  2. Clear All Memory: With the unit off, press and hold the MEMORY button for more than 3 seconds until "CLr" appears, then release. All memory will be erased.
  3. Average of Last 3 Readings: If you take 2 or 3 measurements within 10 minutes, the unit automatically calculates the average. Press MEMORY to view the average reading (indicated by the average icon).

Note: When memory is full (90 readings), the oldest reading is automatically deleted to store a new one.

Care & Maintenance

Proper care ensures accuracy and longevity.

CAUTION! Do not disassemble or attempt to repair the unit. There are no user-serviceable parts inside.

Regulatory Information

This device complies with Part 15 of the FCC Rules. Operation is subject to the condition that it does not cause harmful interference. CE marked, indicating conformity with European health, safety, and environmental protection standards. Manufactured for Omron Healthcare, Inc. For support in the US and Canada, visit www.omronhealthcare.com or call 1-800-634-4350.

Troubleshooting

SymptomPossible CauseCorrective Action
No display when START button pressedDead batteries, loose AC adapterReplace all 4 batteries; ensure AC adapter is firmly connected.
Error Code "E" appearsCuff not attached, air leak, tube kinkedEnsure cuff is plugged in correctly; check tube for kinks; rewrap cuff snugly.
Error Code "Er1" or "Er2"Excessive movement during measurementRemain still and quiet; retake measurement.
Inconsistent readingsIncorrect cuff position, talking/moving, recent activityEnsure cuff is at heart level on bare arm; rest 5 min before measuring; remain still.
Memory not displayingMemory cleared or unit malfunctionCheck if memory was accidentally cleared. If problem persists, contact service.
Irregular Heartbeat symbol flashesIrregular rhythm detectedConsult your physician if this symbol appears frequently.
Cuff inflates too much or too littleIntelliSense calibration issueEnsure cuff is applied correctly. Unit self-calibrates; if problem continues, contact service.

Reset: Remove batteries and AC adapter for 1 minute, then reinstall.

Service: Do not attempt repairs. For service, contact Omron Healthcare at 1-800-634-4350 or visit www.omronhealthcare.com.