Omron Blood Pressure Monitor HEM 711 Manual

The Omron HEM 711 is a fully automatic digital blood pressure monitor for upper arm measurement. It features IntelliSense technology for accurate readings, a large LCD display, one-touch operation, hypertension indicator, irregular heartbeat detector, memory storage for up to 14 readings, date and time display, and automatic power-off. Powered by 4 AA batteries or AC adapter (optional). Below are key sections for safety, features, preparation, measurement, maintenance, and troubleshooting.

1. Features at a Glance 2. Preparation Before Use 3. Taking a Measurement 4. Specifications 5. Controls and Display 6. Memory Function 7. Care & Maintenance 8. Important Notes 9. Regulatory Information 10. Troubleshooting

Features at a Glance

Key components: Main unit with LCD display, Arm cuff with D-ring, AC adapter port (optional), Air tube, Battery compartment, START/STOP button.

FeatureDescription
IntelliSense TechnologyAutomatically inflates to optimal pressure for accurate measurement
Hypertension IndicatorDisplays symbol if systolic/diastolic readings are high
Irregular Heartbeat DetectorDetects and displays symbol for irregular heartbeats during measurement
Memory FunctionStores up to 14 readings with date and time
One-Touch OperationSimple START/STOP button for easy use
Large LCD DisplayShows systolic, diastolic, pulse, and indicators clearly
Automatic Power-OffTurns off after 1 minute of inactivity to save batteries
Cuff Wrapping GuideEnsures proper cuff placement for accurate results
Date & Time DisplayShows when each reading was taken
AC Adapter CompatibleOptional AC adapter for continuous use

Preparation Before Use

Unpack and inspect contents. Ensure all parts are present.

  1. Insert batteries: Open battery compartment, insert 4 AA batteries observing polarity.
  2. Set date and time: Press MEMORY button, use SET button to adjust.
  3. Apply cuff: Wrap cuff on bare upper arm 1-2 cm above elbow.
  4. Position arm: Rest arm on table with cuff at heart level.

WARNING! Do not use on infants, small children, or persons with arrhythmia without consulting physician. Ensure cuff is properly sized.

Taking a Measurement

Remain still and quiet during measurement.

  1. Sit comfortably: Feet flat on floor, back supported, legs uncrossed.
  2. Position cuff: Ensure air tube runs down center of arm.
  3. Press START/STOP: Unit will inflate automatically.
  4. Read results: Systolic, diastolic, and pulse appear on display.
  5. Interpret symbols: Hypertension or irregular heartbeat indicators if present.
  6. Store reading: Automatically saved to memory (up to 14 readings).

CAUTION! Do not move or talk during measurement. Wait at least 3 minutes between readings on same person.

Specifications

Controls and Display

Front Panel: START/STOP button, MEMORY button, SET button.

Display Symbols: Systolic value, Diastolic value, Pulse rate, Hypertension indicator, Irregular heartbeat symbol, Memory number, Date/Time, Battery low indicator, Cuff wrapping guide.

Operation: Press START/STOP to begin measurement. Press MEMORY to review stored readings. Press SET to adjust date/time.

Memory Function

Stores up to 14 readings with date and time.

  1. View memory: Press MEMORY button to cycle through stored readings.
  2. Clear memory: Hold MEMORY button for 3 seconds until display clears.
  3. Memory full: Oldest reading is overwritten when memory is full.

Tip: Record readings in a log book for long-term tracking.

Care & Maintenance

Clean with soft, dry cloth. Do not immerse in water.

  1. Cuff cleaning: Wipe with damp cloth and mild detergent, then dry completely.
  2. Storage: Store in cool, dry place away from direct sunlight.
  3. Battery replacement: Replace when low battery symbol appears.
  4. Air tube: Avoid sharp bends or kinks in tube.

CAUTION! Do not use alcohol, benzene, or abrasive cleaners. Do not disassemble unit.

Important Notes

For accurate measurements:

WARNING! This device is for home use only, not for professional medical diagnosis.

Regulatory Information

FDA Cleared: For over-the-counter home use. CE Marked: Meets European medical device standards. Manufactured by Omron Healthcare Co., Ltd. For support, visit www.omronhealthcare.com or contact 1-800-634-4350.

Troubleshooting

SymptomPossible CauseCorrective Action
No displayDead batteriesReplace all 4 AA batteries.
Error E1 displayedCuff not wrapped properlyRewrap cuff according to guide.
Error E2 displayedMovement during measurementRemain still and retry.
Inconsistent readingsWrong cuff positionEnsure cuff is at heart level.
Memory not workingMemory fullClear memory or overwrite old readings.
Cuff doesn't inflateAir leak or tube blockageCheck connections and tube for kinks.

Reset: Remove batteries for 1 minute to reset unit.

Service: 1-800-634-4350; www.omronhealthcare.com