The Omron 5 Series is a comprehensive line of blood pressure monitors designed for home use, featuring advanced IntelliSense technology for accurate readings, one-touch operation, hypertension detection, irregular heartbeat detection, multi-user memory, and body movement detection. Powered by batteries or AC adapter. Below are key sections for safety, features, setup, operation, maintenance, and troubleshooting.
Key components: Main unit with LCD display, Arm cuff, AC adapter port, Battery compartment, START/STOP button, Memory button.
| Feature | Description |
|---|---|
| IntelliSense Technology | Automatically inflates to optimal pressure for accurate readings |
| One-Touch Operation | Simple START/STOP button for easy use |
| Hypertension Indicator | Displays warning symbol if reading exceeds normal range |
| Irregular Heartbeat Detection | Detects and displays irregular heartbeat during measurement |
| Multi-User Memory | Stores up to 100 readings for two users |
| Body Movement Detection | Alerts if movement is detected during measurement |
| Date & Time Stamp | Records measurement date and time |
| Average Reading | Calculates average of last 3 readings |
| AC Adapter Port | For power connection (adapter sold separately) |
| Battery Compartment | Holds 4 AA batteries |
Read all instructions before use.
WARNING! This device is not a substitute for professional medical diagnosis. Always consult your physician about your blood pressure.
Front Panel: LCD display, START/STOP button, MEMORY button. Side: AC adapter port. Back: Battery compartment, cuff connector.
Cuff Components: Arm tube, metal ring, Velcro fastener, air plug.
LCD Display Symbols: Systolic pressure, Diastolic pressure, Pulse rate, User indicator, Memory full, Battery low, Hypertension indicator, Irregular heartbeat symbol, Date/Time, Average reading.
Prepare device for first use.
CAUTION! Use only recommended batteries. Remove batteries if device will not be used for extended period.
For accurate results, follow these steps carefully.
Tip: Avoid eating, smoking, or exercising 30 minutes before measurement. Rest for 5 minutes before measuring.
Store and review measurements.
Note: Memory full symbol appears when 100 readings stored for a user.
Interpret your blood pressure readings.
| Category | Systolic (mmHg) | Diastolic (mmHg) |
|---|---|---|
| Normal | Less than 120 | and Less than 80 |
| Elevated | 120-129 | and Less than 80 |
| Hypertension Stage 1 | 130-139 | or 80-89 |
| Hypertension Stage 2 | 140 or higher | or 90 or higher |
| Hypertensive Crisis | Higher than 180 | and/or Higher than 120 |
Symbol Indicators: Hypertension symbol appears if reading exceeds 135/85 mmHg. Irregular heartbeat symbol appears if detected during measurement.
Proper care ensures accurate measurements and long device life.
CAUTION! Never immerse device in water. Do not use near flammable gases.
FDA Cleared: This device complies with FDA standards for home blood pressure monitors. CE Mark: Meets European medical device directives. Manufactured by Omron Healthcare, Inc. For support, visit www.omronhealthcare.com or contact 1-800-634-4350.
| Symptom | Possible Cause | Corrective Action |
|---|---|---|
| No display | Dead batteries | Replace all 4 AA batteries |
| Error E1 displayed | Cuff not connected properly | Check cuff connection; reinsert plug |
| Error E2 displayed | Cuff pressure error | Check for cuff leaks; replace if damaged |
| Error E3 displayed | Measurement error | Remain still; retake measurement |
| Inconsistent readings | Incorrect cuff position | Ensure cuff at heart level; retake measurement |
| Memory not working | Memory full | Clear memory or switch to other user |
| Device turns off during use | Low batteries | Replace batteries or use AC adapter |
Reset: Remove batteries for 1 minute to reset device.
Service: 1-800-634-4350; www.omronhealthcare.com